Diamond ProLYTE
 
 
 
 
 
 
 

 
SPECIFICATIONS:
  Sample
Whole Blood, Serum, Plasma, or Urine

Data Storage
125 patient results
20 normal QC results
40 abnomal QC results
Output
16x2 place alphanumeric display
40 column printer
Serial port (RS-232)
AutoSampler port
       
  Sample Size
100 µl Whole Blood, Serum or Plasma
60 µl Capillary Sample
400 µl diluted(1:10) Urine
Calibration
Automatic or On Demand
Ambient Conditions
15-32°C (60-90°F),<85% humidity
       
  Detection Range
Blood Na+: 20-200 mmol⁄L
            K+:  0.2-40 mmol⁄L
             Cl:  25-200 mmol⁄L

Urine Na+: 25-1000 mmol⁄L(undiluted)
           K+:  1.0-500 mmol⁄L(undiluted)
           Cl:  25-500 mmol⁄L(undiluted)

Analysis Time
55 sec (Blood)
90 sec (Urine)
Reproducibility
Typical Within Run (n=20)
Blood, Serum, Plasma
Na+: CV≤1% (80-200 mmol⁄L)
K+: CV≤2% (1-10 mmol⁄L)
Cl: CV≤2% (80-200 mmol⁄L)

Urine
Na+: CV≤5% (25-1000 mmol⁄L)
K+: CV≤5% (1-500 mmol⁄L)
Cl: CV≤5% (25-1000 mmol⁄L)
Power
100-115 ˜ VAC 50-60 Hz, 0.8 A or
220˜VAC 50⁄60 HZ, 0.4 A
(Factory set)



Size and Weight
9.5" (24cm) W x 16.5" (42 cm) H x
8.0" (20cm) D, 13 lbs. (5.8 kg)
This analyzer is produced and distributed by Diamond Diagnostics Inc. There is no affiliation, connection, sponsorship or association between Diamond Diagnostics, Inc. and Medica Corporation.